For ovens up to 700°C several standard viewing port and window options are available. For non-gas tight ovens cable ports with 6, 36 or 48 mm diameter may be fitted in the side wall and windows in the oven door.
For most standard chamber furnaces up to 1800°C viewing ports fitted on the back wall or furnace roof are available on request.
For tube furnaces up to 1100°C quartz glass windows can be fitted on request.
For most chamber furnaces up to 1800°C viewing ports made be fitted on the furnace door either with or without quartz glass windows (depending on the size). Thermocouple may also be fitted through the rear wall of most standard chamber furnaces. Calibration ports are also available.
The sliding thermocouple is a standard option for pyrometer controlled vacuum chamber furnaces and is located in the hot area of the furnace to control and monitor accurately the temperature up to 1500 °C. At temperatures above 1500 °C, the thermocouple automatically slides out of the hot zone and the pyrometer takes over temperature measurement.
Pyrometers are used for vacuum chamber furnaces above 2200°C for temperature control and are fitted as standard. In other furnaces, pyrometers can be fitted on request.